Picarro G2508 CO2 ,CH4 ,N2O ,NH3 和H2O分析儀同時(shí)輸出自然態(tài)氣體參數(shù)和當(dāng)時(shí)水汽濃度下的摩爾干值,配備ChemDetect干擾檢測標(biāo)記的軟件。無論是青藏高原的土壤通量測量,還是海洋船只上的原位測量,事實(shí)證明Picarro出色的野外應(yīng)用能力。
技術(shù)原理第四代波長掃描光腔衰蕩光譜技術(shù)(WS-CRDS)
技術(shù)參數(shù)
G2508 痕量氣體分析儀 | |||||
性能指標(biāo) | CO2 | CH4 | N2O | NH3 | H2O |
精度 (初始精, 1σ) | <600ppb +0.05%讀數(shù) | <10ppb +0.05%讀數(shù) | <25ppb +0.05%讀數(shù) | <5ppb +0.05%讀數(shù) | 500ppm |
精度 (1min, 1σ) | <300ppb +0.05%讀數(shù) | <7ppb +0.05%讀數(shù) | <10ppb +0.05%讀數(shù) | <3ppb +0.05%讀數(shù) | 250ppm |
精度 (5min, 1σ) | <200ppb +0.05%讀數(shù) | <5ppb +0.05%讀數(shù) | <5ppb +0.05%讀數(shù) | <1ppb +0.05%讀數(shù) | 100ppm |
確保精度范圍 | 380~5000ppm | 1.5~12ppm | 0.3~200ppm | 0~300ppb | 0~3% |
測量范圍 | 0.02~2% | 0.5~15ppm | 0~400ppm | 0~2ppm | 0~7% |
測量速率 | <8s | <8s | <8s | <8s | <8s |
響應(yīng)時(shí)間(10-90%/90-10%) | ~8s | ~8s | ~8s | 見備注 | ~8s |
摩爾量(干值) | Yes | Yes | Yes | No | N/A |
備注:由于實(shí)驗(yàn)裝置表面吸附的限制,儀器對H2O和NH3的響應(yīng)受到影響。但是儀器可以準(zhǔn)確測量在腔室內(nèi)測量來自土壤呼吸的H2O和NH3,并計(jì)算準(zhǔn)確的土壤呼吸通量。
取樣溫度 | -10 ~ 45 ℃ | 溫度控制精度 | ±0.005℃ |
環(huán)境濕度 | <99% R.H,無冷凝 | 壓力控制精度 | ±0.0002atm |
取樣壓力 | 300-1000Torr(40-133kPa) | 閉路循環(huán)系統(tǒng) | 兼容Picarro A0702 |
取樣流速 | ~230mL/min@760Torr,無需過濾 | 重量 | 27.2Kg(包括外置泵) |
功耗 | 100-240 VAC,啟動(dòng)時(shí)<260W;穩(wěn)定后:分析儀110 W,泵35 W | 尺寸 | 分析儀:43.18x17.78x 44.57cm外置泵:19 x 10.2 x 28 cm |
Picarro CRDS技術(shù)采用了極窄的光譜區(qū)域,與其它光譜技術(shù)相比,**限度的減少了其它種類氣體的干擾。但是,現(xiàn)實(shí)世界的樣品干擾還是可能存在的。Picarro提供光譜干擾檢測軟件,測試并給出以下氣體種類對N2O的干擾與否數(shù)據(jù)。
CO2 | 無干擾:自動(dòng)校準(zhǔn)至20000ppm | 乙烷 | 0.2 ppb N2O / ppm C2H6測試至120ppm |
CH4 | 無干擾:自動(dòng)校準(zhǔn)至200ppm | 乙烯 | 0.5 ppb N2O / ppm C2H6測試至16ppm |
NH3 | 無干擾:自動(dòng)校準(zhǔn)至2ppm | 乙炔 | 不供使用 |
ChemDetect™ | 檢測和標(biāo)記光譜干擾的數(shù)據(jù) | 背景氣 | 大氣應(yīng)用,非高度變化或富集的N2, O2, H2, He |